  1. i) Patient and operator safety and system performance should not be adversely affected by fluid spills.
  2. ii) If the unit is affected, it should fail safely.

 

  1. c) Overcurrent Protection
  2. i) Loss of power to other equipment on the same branch circuit due to internal equipment faults should be

prevented by using fuses or circuit breakers that are clearly labeled and easy to replace or reset.

  1. ii) If fuses are used, a spare fuse should be provided in a labeled holder located next to the main fuse holder.

Permanent markings near each fuse holder should indicate fuse ratings.

 

  1. d) Line Voltage Variation
  2. i) The unit should operate satisfactorily at line voltages from -12.5% to +8%

of the nominal line voltage of 220Volts.

  1. ii) The unit should not be damaged by voltages from -21% to +12.5% of the nominal line voltage of 220Volts.

 

 

 

  1. e) Electromagnetic Interference (EMI)
  2. i) The unit’s performance should not be affected by EMI radiated or conducted through the power lines from another

device.

  1. ii) If the unit is affected, it should fail safely.

HICLOVER is growing brand for environmental protection field, and market share with most of Africa, Middle East, Southeast Asia countries and part of North America, Europe territory. We are trusted partner for governmental organizations, non-profit organizations, international contractors, logistics organizations, military, pet cremation business owners, etc. We have export experience more than 40 countries, including war zone like Iraq, Afghanistan, Somalia, South Sudan. Mobile: +86-13813931455(WhatsApp) Website: www.hiclover.com Email: [email protected] Email: [email protected]